Positive Displacement Calibrator Liquid – PDCL 10

PDCL 10 – Primary Standard Calibrator Liquid
The Positive Displacement Calibrator Liquid (PDCL 10) provides calibration for meters 1/2" or smaller. The test section is large enough for multiple flowmeter calibrations, up to 4 depending on stack length limitations. The PDCL 10 comes with the same features as the larger models. Of particular importance is the three valve fluid flow rate control valves – course, medium and fine – for ease of obtaining a precession flow rate. Combining the PDCL 10 with the PDCL 400 will cover meters from 3/8" to 2–1/2" or 0.01 to 400 GPM.
This feature pack Primary Standard Liquid Flow Calibrator is easy to maintain and will last 50+ years in service. The PDCL 10 conforms to the guidelines of NIST for Round Robin Testing.
Model PDCL–10 Abbreviated Specification:
- Fluid: Hydrocarbons & Water
- Range: 0.001 to 10 GPM
- Viscosity Range: 0.5 to 1000 Centistokes
- Uncertainty: ±0.05% of reading
- Ambient Operating Temperature: 60 to 90 Degrees F.
- Fluid Operating Temperature: 40 to 140 Degrees F.
- Operating Pressure up to: 120 psig
For a complete specification view the PDCL Data Bulletin